50ma mobile x-ray equipment
- Payment Type:
- T/T
Your message must be between 20 to 2000 characters
Contact NowPlace of Origin: | China |
---|---|
Payment Type: | T/T |
50ma mobile x-ray equipment
General Description:
1. This medical radiography x-ray equipment is a full wave rectification photographic medical diagnosis machine whose tube head can take photos. It is suitable for X ray photographing in wards and operation rooms in all hospitals as well as special hospitals, especially in orthopaedics-wards. It can also used for X ray photographing in the family sickrooms.
2. Due to its adoption of high voltage silicon stack full-wave rectification technology, the efficiency of the X ray tube used in the machine is heightened.
3. This machine has a kV, mA and s capacity interlocking control system, and an overload indicating so as to prevent the tube head of the machine from damaging due to the accident operation.
4. Either the wired manual exposure or the remote control exposure may be used by the technician at his selection during photographing. In the course of exposure, there will appear a sound and light indication for the convenience of patients and technicians.
5. This machine is equipped with an electric beam limiting device, on which there is an indicator indicating the ray beam aiming at the center of the X ray film, the position of the focus center of the patient against the center of the X ray film can thus be determined easily with reference to the indication.
6.This machine is also provided with a tube head cantilever moving up and down. It is easy to set the tube head of the machine to required angle and height for the convenience of photographing. If it is necessary to move the whole machine, lower the machine tube head cantilever to its lowest position, and lock it up for the convenience and safety of machine movement.
7.The total weight of the x-ray equipment is 90 kg, and easy to be moved with a light handcart (65 cm in width) through the corridor in the ward of hospital and between hospital beds.
Main purpose and features:
1. Used for X ray radiography in wards, especially in orthopaedic wards.
2. Single focus point, high voltage silicon stack full-wave rectification.
3. kV, mA and s capacity interlocking control system, overload indicating system.
4. Wired manual exposure & remote control exposure
5. Sound and light indication during the course of exposure.
Main Technical Data
1. Power requirements
Voltage:180-240V Frequency:50Hz; Internal resistance: ≤1Ω Power capacity: 5k VA ; Current (Max., instantaneous) <20A
2. Timer: 0.08s to 6.3s, in 20 shifts, electronic type
3. Maximum distance from the focus of the X-ray tube to the ground ≥1820 mm
4. Minimum distance from the focus of the X-ray tube to the ground ≤550 mm
5. The maximum exposure field is 350×350 mm if the distance between the
focus of the Xray tube and the film is 650 mm, and the maximum exposure
field is 450×450 mm if the distance between the focus and the film 1000 mm.
6. The remote control exposure distance ≥ 6m.
7. X-ray tube focus:2.3×2.3
8. The required pulling force to move the whole machine ≤50N (5kg)
9. Net weight of the unit:90kg
Maximum ratings